Analytical Principle
Infrared absorption, at both 3 and 9μm (microns) – on every sample
Range
0.000 to 0.600 %BAC [g/210 Liters]; 0.000 to3.000 mg/liter
Dimensions
4.5” long x 10” wide x 8.4” deep
Weight
17 lbs [instrument only; just over 20 lbs with optional gas delivery system
Power Ratings
AC: 90 – 264 VAC, 47 – 63Hz, 1amp max.
Vehicle DC: 10 – 16 VDC, 6 amps max.
Microprocessor
Intel compatible x86 based processor @ 30 MHz with a separate DSP processor for signal conditioning
Memory
2 MB for data storage [expandable to 8 MB]; 1 MB FLASH/SRAM for code/fast access; remote program updates
Communications
Up to three RS232 serial ports, one external printer port, internal modem, Ethernet, USB host, USB device, Bluetooth
Calibration Checks
Wet bath simulator or dry gas
CE Certification
Complies with IEC 1010-1 (1990, 92, 95)
OIML
Complies with OIML R 126 [1998 (E)]
RFI Certification
FCC Part 15 Class B and BS EN 61000-6-1,2,3,4,5,6,8,11
Vibration and Shock
To IEC 68-29Eb,6Fc,64Fh,27Ea
Software Configuration
Various software settings may be configured by the instrument Manager through the keyboard [via a security password] / through an external PC using COBRA